Product Description:
MOX1 MOX-1 Measuring O2 Oxygen Concentration Electrochemical Gas Sensor
Features:
N.B. The specification is based on measurements made with cylinder gases using a flow rate of 100
mls min-1. Conditions at 20°C, 50%RH, and 1013mBar unless otherwise noted.
Intended Use
These sensors are designed to be used to monitor the partial pressure of oxygen in anaesthesia, critical care,
incubators and general Oxygen monitors.
Stabilisation time
Allow at least 15 minutes to stabilise in instrument before calibration.
Cleaning and Sterilisation
In case of contamination the sensor may be cleaned with distilled water and allowed to dry naturally. The sensor is
not suitable for sterilisation by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide
Calibration Interval
These sensors are designed to have minimal drift over their useful lifetime however for maximun accuracy they should
be calibrated in 100% Oxygen before use.
Cross-sensitivity
Test Gas Error (%O2)
50% He/50% O2 <1%
80% N2 O/20% O2 +1 to +1.5%
4% Halothane/28.8% O2 /67.2% N2O +1.5% to 2%
5% Sevoflurane/28.5% O2 / 66.5% N2O +1 to +1.5%
5% Enflurane/28.5% O2 /66.5% N2O +1.2 to + 1.8%
5% Isoflurane/28.5% O2 /66.5% N2O +1.2 to 1.8%
5% CO2 / 28.5% O2 /66.5% N2O <1%
Sensor Connections Suggested Mating Parts
MOX-1 M16 thread, 3-pin molex header
(MOLEX 22-29-2031)
MOLEX 22-01-2035 (3-way housing)
MOLEX 08-56-0110 (crimp terminals)
MOX-2 M16 thread, 0.141”/3.5mm mini
phone jack
SWITCHCRAFT 750 mini phone jack
MOX-3 M16 thread, UK data socket
AMP P/N 520257-2
MOX-4 M16 thread, UK data socket
AMP P/N 520257-2
Specifications:
Measurement range | 0-1500mBar O2 |
Response Time (T90) | <15 s (air to 100% O2) |
Baseline Offset | <200 μV |
Linearity | Linear 0% to 100% O2 |