Product Description
This system uses tap water for direct connection to realize the unified supply of various departments, and is safely connected with various brands of automatic disinfection machines, cleaning machines and sterilization equipment.
In your lab,what applications require pure water?
1.feeding glassware washing machines and autoclaves
2.filling water baths and incubators
3.making reagents
4.Analytical work such as HPLC,ICP-MS,GC-MS
5.tissue culture media and buffes
Technical Parameter
Type | Comprehensive Analytical Type |
Model | OK-EP20VF |
AC Voltage | 110/220V,50/60Hz |
Flow Rate | 1.5-2L/min |
Output | 20L/h |
Feed Water Requirements | Tap water, temperature 10~35℃,pressure 0.15-0.35Mpa;TDS:≤200mg/L |
Water Quality | |
Type 2 Pure Water | Conductivity:2-10μs/cm@25℃ |
Type 1 Ultra-pure Water |
Resistivity:18.25MΩ/cm@25℃, TOC:1-3ppb, Bacteria: <1CFU/ml, Particle: (>0.2μm) <1pc/ml |
If UV or Not | Yes.UV lamp to get rid of bacterial spores |
If Ultrafiltration or Not | Yes.Ultrafiltration to pyrogen removal |
Application |
Molecular biology, life science, gene research, tissue culture, test tube baby, bioengineering, animal and plant cell culture, DNA, monoclonal antibody research, etc. |
Dimension(L*W*H-mm) |
410*300*500 |
Weight(kg) |
30 |
Consumable Cartridge |
Including a standard cartridge PP filter, activated carbon filter, |
Maintenance and Tips |
Generally,change it every 6months(PP filter,actived carbon filter)and replace it when resistivity ≤ 10MΩ.CM(iox exchange column) |
Laboratory Water Purifier
Type 1 Ultra Pure Water
• Applications:
• Analysis of organic and inorganic traces
• Chromatography
• Mass spectrometry
• UV-visible spectrophotometry
• Microbiology
Type 2 Pure Water
• Applications:
• Media preparation
• Preparation of buffers
• Washing and rinsing of glassware
• Preparation of chemical reagents
• Ultrafiltration equipment washing
Ordering Information