Edan IM60 iM50 Patient Monitor Battery Connector Board 02.02.114507-11 21.53.114506-1.0
► Brief Introduction
This monitor can run on battery power, which ensures its uninterrupted operation even when AC power supply is interrupted. The battery recharges whenever the monitor is connected to the AC power source. During monitoring, if the AC power is interrupted, the monitor will take power from the internal battery. If the monitor is powered by battery, the monitor will switch off automatically before the battery is completely depleted.
► Note
- The service life of the battery depends on the service frequency and time. The service life of the battery is about three years if the battery is well maintained and stored. The service life of the battery may shorten if it is used inappropriately. If the battery life is exhausted and not replaced in time, it may cause damage or heat to the device.
- Periodic checks on the battery performance are required. Change the battery if necessary.
- Do not connect the positive (+) and negative (-) terminals with metal objects, and do not put the battery together with metal objects, which can result in short circuits.
- Do not unplug the battery when the monitor is working.
- Do not heat or throw the battery into a fire.
- Do not use, leave the battery close to fire or other places where temperature may be above 60°C.
- Do not immerse, throw, or wet the battery in water/seawater.
- Do not destroy the battery: do not pierce the battery with a sharp object such as a needle; do not hit with a hammer, step on or throw or drop to cause strong shock; do not disassemble or modify the battery.
- Do not solder the leading wire and the battery terminal directly.
- If liquid leaking from the battery gets into your eyes, do not rub your eyes. Wash them well with clean water and go to see a doctor immediately. If liquid leaks of the batterysplash onto your skin or clothes, wash well with fresh water immediately.
- Keep away from fire immediately when leakage or foul odor is detected.
- Stop using the battery if abnormal heat, odor, discoloration, deformation or abnormal condition is detected during use, charge, or storage. Keep it away from the monitor.
► Specification
Operating Time | 2.1Ah | 150 min or longer |
4.2Ah | 300 min or longer |
Condition | At 25°C, with (a) new fully charged battery/batteries, continuous SpO2 measurement and NIBP automatic measurement mode at interval of 15 minutes, ECG/TEMP module connected, recording at interval of 10 minutes, brightness set to “1” |
Charge Time | 2.1Ah | 200 min or shorter |
4.2Ah | 360 min or shorter |
Condition | Monitor is on or in standby mode. |
► Battery Status
1.
Remaining battery power: 76%~100%.
2.
Remaining battery power: 51%~75%
3.
Remaining battery power: 26%~50%
4.
Remaining battery power: 4%~25%
5.
Batteries are almost depleted and need to recharge immediately.
6.
No battery is installed
7.
Battery malfunction
Please kindly contact us, if you are interested in Edan IM60 iM50 Patient Monitor Battery Connector Board 02.02.114507-11 21.53.114506-1.0 , Thank you!