Metrax Primedic Multifunction Defibrillator Electrodes 97796 SavePads For AED defibrillator 96389
►Brief Introduction
The PRIMEDIC™ HeartSave PAD is intended to be used by non-experts and trained first-aiders. It may only be used on patients who are unconscious and who are not breathing.
The main application is to carry out transthoracic defibrillation in asynchronous mode. The decision about the necessity of applying a shock is carried out automatically by the device.
The operator is guided by spoken instructions and clear written and pictorial information. After switching on the device, the patient is connected to it using the enclosed adhesive electrodes. After this, automatic rhythm analysis is carried out by the device. Only if a rhythm is detected by the device as being ventricular fibrillation (=shockable) does it suggest treatment with a high energy shock. All other rhythms are classified as not requiring a shock. The time from the start of analysis until the shock is applied is less than 30s.
The energy levels of the first, second and third shock are predefined by the maximum set points for the electric current 20A, 25A and 30A, as well as by the capacitor voltage which depends on the patient impedance.
1. Plug
2. Defibrillation electrodes with protective film
► Opening SavePads and positioning electrodes
Procedure:
1. Open the defibrillation electrodes bag by tearing open the protective cover along the tear strip.
2.Remove the protective film (1) from one of the electrodes (2) and then immediately place the electrode on the position you had ascertained previously. Refer to label stuck on the back of the electrodes.
3.Proceed to remove the protective film from the second electrode and place it in its position.
4.Smooth the electrodes onto the patient ensuring there are no air bubbles under the electrodes! Display < Apply electrodes to patients bare chest > must go out.
► Output power in AUTO mode at:
Patient impedance | 1st stage | 2nd stage | 3rd stage |
25 Ohm | 143 J | 201 J | 277 J |
50 Ohm | 281 J | 350 J | 360 J |
75 Ohm | 348 J | 360 J | 360 J |
100 Ohm | 344 J | 343 J | 343 J |
125 Ohm | 314 J | 316 J | 317 J |
150 Ohm | 290 J | 293 J | 293 J |
175 Ohm | 269 J | 272 J | 272 J |
► Delivery Issues
As a general rule, all the orders are going to be shipped within 3 to 5 working days and delivered on time, we will try our best to process the orders and ship them out as soon as possible.
To make sure the business goes smoothly, please pay special attention to it and carefully arrange your time of placing orders according to your personal needs. If you are in urgent need of your item, please be sure to confirm with us before placing your orders. If you have any questions such as your order is taking much more time than expected or you need to know when will your order arrive, you can contact us at anytime. We will reply you as soon as possible.