PHiliphh MP5 patient monitor ECG SPO2 NIBP parameter connector panel Measurement Block M8105-60040
► Brief Introduction
This is the area of the PHiliphh MP5 patient monitor where various monitoring cables and sensors are connected. It typically includes designated input ports for ECG, SpO2, NIBP, and other parameters. The panel ensures that the monitor receives accurate data from the sensors and displays the corresponding vital sign information on the screen.
► Features
ECG (Electrocardiogram): ECG monitoring involves measuring the electrical activity of the heart to assess its rhythm and function. Electrodes are placed on the patient's body, and the ECG cable is connected to the ECG input on the parameter connector panel. The ECG waveform and heart rate are displayed on the monitor.
SpO2 (Oxygen Saturation): SpO2 monitoring measures the oxygen saturation level in the patient's blood. A sensor is typically placed on a finger or another part of the body, and the SpO2 cable is connected to the SpO2 input on the parameter connector panel. The patient's oxygen saturation level is displayed on the monitor.
NIBP (Non-Invasive Blood Pressure): NIBP monitoring measures the patient's blood pressure without the need for invasive procedures. A blood pressure cuff is placed on the patient's arm, and the NIBP tubing is connected to the NIBP input on the parameter connector panel. The monitor periodically inflates the cuff to measure the patient's blood pressure, and the results are displayed on the monitor.
► Removing the Measurement Block
1. Pull off the Measurement Block Cover by flipping it upwards from the bottom as shown below.
2. Unplug the ribbon connector from the Front End 2 measurement block to the measurement board.
3. Unscrew the screws from the Front End 2 measurement board.
4. Lift off the Front End 2 measurement block.
5. Replace the Front End 2 measurement block, if necessary.
6. Exchange Front End 2 housings contain a CO2 inlet which may not fit with your Front End 2 Measurement block cover. This may be the case if your monitor was originally not equipped with Microstream CO2. If you have the old version without the CO2 inlet, separate the Front end 2 measurement block housing of the exchange part as shown below.
7. Unplug the end of the NBP tubing and remove the tubing from its holders.
8. Unplug the connectors from the Front End 1 measurement block.
9. Unscrew the screws securing the Front End 1 measurement block and remove the measurement block.
10. For reassembly, perform the above steps in reverse order.
► Specification
Operating temperature | 0 to 40ºC (32 to 104ºF) |
Storage temperature | -20 to 60ºC (-4 to 140ºF) |
Operating humidity | 20 to 85% RH (Non-condensing) |
Storage humidity | 5 to 85% RH (Non-condensing) |
Operating Altitude | Up to 4000m (13123.36 ft.) |
Storage Altitude | Up to 12000m (39370,08 ft.) |
Thank you for your time! If you are interested in PHiliphh MP5 patient monitor ECG SPO2 NIBP parameter connector panel Measurement Block M8105-60040, welcome to place your orders. If you want to explore more details please let us know.