Hospital Equipment Parts GE Dash4000 Patient Monitor Encoder Keypad Board Assembly Function Keys
► Brief Introduction
This keypad assembly containing multiple buttons.This function keys control plus 5 hard keys: Power, GraPHiliphhilip Go/Stop, NBP Go/Stop, Zero All, and Silence Alarm/Admit.
The following keys are on the monitor’s control panel:
Power | Press once to turn the monitor on and off. |
GraPHiliphhilip Go/Stop | Press once to print patient data or an information window. Press again to stop the print job. |
NBP Go/Stop | Press once to start one noninvasive blood pressure measurement. Press again to stop the measurement. |
Zero All | Press once to zero all invasive pressures lines which are open to the atmosPHiliphhilipere. |
Silence Alarm/Admit | Press Silence Alarm/Admit to quickly admit a patient to the monitor when the monitor is in the discharged state. |
Power:
Press the Power key to begin the normal operating mode. Normal mode provides all functional capabilities of the monitor, including vital signs monitoring, network communication and battery charging. During the power-up sequence, a splash screen displays and the control panel and alarm light indicators illuminate. After approximately 20 seconds, the Main Menu displays.
NBP:
Automatic non-invasive blood pressure monitoring uses the oscillometric measurement method. The oscillometric method uses a sensitive transducer which measures cuff pressure and minute pressure oscillations within the cuff to determine the mean pressure and calculate the systolic and diastolic pressures. A parameter window display updates every second. There is no waveform.
► Replace Keypad Assembly
► Temperature