Electric defibrillation is the process of striking the heart with a certain amount of current, in order to
Methods for terminating ventricular fibrillation. The use of defibrillators involves coating the electrode plates
Good conductive paste, then place saline gauze on the patient's chest wall, press the charger button, and you can hear a continuous charging sound. Place the electrode plate on the patient's chest wall, with the negative electrode placed at the junction of the left fifth intercostal space and the midaxillary line, at the apex of the heart, and the positive electrode placed at the second intercostal space on the right edge of the sternum, at the bottom of the heart. The operator should stretch their arms straight, fix the electrode plate, keep their body away from the edge of the bed, and keep people around them away from the bed to avoid electrical damage. The operator needs to simultaneously press the discharge button with both hands to defibrillate. What is the selection of defibrillation energy
Choose 360 joules for unidirectional waves and 150-290 joules for bidirectional waves
Ears.
1. Select to enter the "power on" or "defibrillation" mode;
2. Select "30J" and press the "Charge" button;
3. Place the left and right defibrillation handles on both sides of the machine base, and continuously press the "SHOCK" button on both handles to discharge. The screen prompt "test normal" indicates that the handle buttons and discharge circuit are working properly;
(Once a month) Select "200J" and press the "Charge" button. Wait for the screen to prompt "Charge Complete" and hold for 1 minute for the host to discharge the battery internally (during which there are no fault code prompts on the screen), indicating that the high-voltage circuit is working properly;
Cleaning of external defibrillation handle:
First, remove all dirt (tissues, body fluids, etc.) from the surface and thoroughly wipe it clean with a cloth dipped in water. Then, use one of the following products for cleaning:
Isopropanol (70% aqueous solution).
Flexible soapy water.
Chlorine bleach (containing 6% sodium hypochlorite), prepared as a 3% aqueous solution.
Cleaning solution/wipes containing flexible isopropanol and concentrated chlorine bleach.
Summarize the self checking methods of defibrillators;
Attention points:
1. Self inspection requires disconnecting the AC power cord;
2. Not used normally, need to plug in the power supply to ensure full power and standby;
3. Clean the conductive paste or other stains on the electrode plate in a timely manner after use, ensuring that the surface of the electrode plate and the embedded pediatric electrode plate is clean and free of stains;
4. Do not confuse the conductive paste used for defibrillators with ultrasound coupling agents or place the ultrasound coupling agents on the defibrillator trolley;
Type: | Nihon Kohden TEC-5602 Defibrillator External Paddle ND-560V |
Warranty period: | 90 Days |
Shipping Date: | 3-5 Working Days |
MODEL: | TEC-5602 |
Storage temperature: | -20-70°c |
MOQ: | 1 PCS |
COLOR: | Orange |
Delivery Date: | 3 - 5 Working Days |
Package | Standard export packaging |
The company has been professionally operating various brands of original medical monitoring instruments since 2015. The company has many professional senior engineers and has sufficient ability to undertake various medical equipment maintenance work. Covering motherboards, sensors, probes, modules, connectors, shells, etc. for various medical monitoring equipment.
We default to express delivery by DHL,Fedex,etc.We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.