Oxygen sensor calibration test
Test the oxygen sensor calibration as follows:
1. Connect the ventilator oxygen tubing to a known source of 100% O2 (e.g., a medical oxygen cylinder). Press the 100% O2/CAL button for 2
minutes or the INCREASE O2 button for 2 minutes to calibrate the oxygen sensor. Once the button light turns off, proceed to the next step.
2. Connect the ventilator oxygen tubing to another known source of 100% O2 (e.g., a second medical oxygen cylinder).
3. Set %O2 to the following values and wait one minute for each monitored value to stabilize: 21% 40% 90%
4. Observe the upper display to ensure that the O2 value (% O2 delivered) is within 3% of each setting within one minute of selecting each
setting.
Oxygen Sensor Calibration Button: Older ventilators use the 100% O2/CAL 2 minutes button, while newer ventilators use the INCREASE O2 2
minutes button. Provides 100% oxygen (if available) for two minutes and calibrates the oxygen sensor. The green light on this button will
illuminate and a message (100% O2 Cal in Progress) will appear on the lower touchscreen to indicate 100% O2 delivery is active. If you press the
O2 button again, the system will restart the two-minute delivery interval. Press CANCEL to stop calibration. For information on calibrating the
oxygen sensor, see page TR 15-5.
Alarm Messages
If you see this O2 Sensor message, it means that background checks have identified a problem with the oxygen sensor (sensor error or not calibrated). Patient ventilation will not be affected. And you can:
• Press 100% O2 CAL or O2 INCREASE for 2 minutes to recalibrate the oxygen sensor.
• Disable the oxygen sensor.
• Replace the oxygen sensor.
Specification of 10097559 PSR-11-75-KE13PB oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine
Product Name | PSR-11-75-KE13PB 10097559 oxygen sensor for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine
|
Brand | Covidien |
Part No |
PSR-11-75-KE13PB 10097559
|
Application | PB760 PB840 PB980 Puritan Bennett Ventilator Machine
|
MOQ | One piece |
Condition | Original /Compatible new |
Warranty | 90 days |
Photo
Below photo is PSR-11-75-KE13PB 10097559 oxygen sensor ,which can be used for Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine .For more details ,please feel free to contact us .
Thank you so much for your attention .If you're interested in 10097559 PSR-11-75-KE13PB Oxygen Sensor For Covidien PB760 PB840 PB980 Puritan Bennett Ventilator Machine ,feel free to contact us .Thank you .