TMS-1024I 12V20W KYOTO SP2057 biochemical analyzer light source bulb
TN-86226
Generally, when a light bulb needs to be replaced or malfunctions, the KYOTO SP2057 biochemical analyzer will provide a prompt, such as bulb failure 419, malfunction of the photometric measurement system (triggering sensor) 420, malfunction of the photometric measurement system (ADC/lamp) 421, malfunction of the photometric measurement system (trigger) 426, failure of the reaction cup integrity check (A system) 443, bulb warning 444, malfunction of the light bulb 445, when the ADC value is higher than the maximum range 626 during inspection, the above prompt indicates the need to replace the light source bulb.
1. First, wait for the instrument to shut down normally, at least 5 minutes, to allow the light source to cool sufficiently. Open the instrument cover door and remove the protective cover plate from the photometer.
2. Gently remove the transparent cover from the junction box, use a Phillips screwdriver to loosen the two fixing screws, and remove the power cord of the light source lamp from the junction box.
3. Loosen the fixing screws of the light source mounting plate and remove the light source and mounting plate together.
4. Remove the light source lamp from the installation board. Install the light source lamp onto the mounting plate (be careful not to touch the glass part of the bulb with your hands). Install the light source mounting plate into the slot, gently push the mounting plate backwards during installation to ensure that it is installed in place, and tighten the fixing screws
5. Wrap up the excess power cord and connect it to the power box without distinguishing between positive and negative poles.
6. Finally, install the transparent cover of the junction box, cover it with a protective cover, and close the instrument cover door.
Type: | TMS-1024I 12V20W KYOTO SP2057 biochemical analyzer light source bulb TN-86226 |
Warranty period: | 90 Days |
Shipping Date: | 3-5 Working Days |
MODLE: | TMS-1024I 12V 20W bulb TN-86226 |
Storage temperature: | -20-70°c |
MOQ: | 1 PCS |
COLOR: | Metal and glass |
Delivery Date: | 3 - 5 Working Days |
Package | Carton |
The company has been professionally operating various brands of original medical monitoring instruments since 2015. The company has many professional senior engineers and has sufficient ability to undertake various medical equipment maintenance work. Covering motherboards, sensors, probes, modules, connectors, shells, etc. for various medical monitoring equipment.
We default to express delivery by DHL,Fedex, We will contact you as soon as possible when sending express delivery. If you have any questions after receiving the express delivery, please feel free to contact us at any time.