Specifications
Brand Name :
Daruifuno
Model Number :
ADS2-U3A-CD7
Certification :
CE
Place of Origin :
China
MOQ :
Negotiable
Price :
Negotiable
Product Name :
Chlorine Dioxide Analyzer
Product Model :
ADS2-U3A-CD7
Software Version :
DRFN Chlorine Dioxide Analysis software V1.0
Display Range :
0.00-200.00mg/L
Protection Grade :
IP66
Size :
144*144*120mm
Power Supply :
100~240V AC
Weight :
800g
Power Consumption :
12W
Mounting Method :
Wall mount, pipe clamping, panel
Description

11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane

Introduction

When chlorine dioxide is used for disinfection, it is best to use it during production on site. Chlorine dioxide produces non-toxic substances after sterilization, which does not pollute the environmental water body.

The ADS2-U3A-CD7 chlorine dioxide analyzer is a single-parameter instrument. The chlorine dioxide controller is connected to the DRFN sensor for measuring chlorine dioxide with a surfactant-resistant membrane at 1~11 pH.

Advantages Of Chlorine Dioxide Analyzer

1. Menu password can be set to facilitate hierarchical management;

2. All the configuration entered by the user is kept indefinitely, even if there is no power, the data will not be lost.

3. 128*64 black and white graphic dot matrix screen, and the display backlight provides always on and energy-saving mode selection.

4. It has time and historical data recording functions to meet higher data management requirements.

5. Abundant output meets monitoring, control, and remote data transmission at the same time.

6. Two settable 0/4~20mA active current loop outputs, and each current output allows the user to specify what measurement value it represents.

7. Two settable SPST relays. The setting objects include alarms for the high and low values of the measurement parameters, the control of the cleaning cycle of the sensor. The warning of the working environment temperature of the instrument.

Applications Of Chlorine Dioxide Analyzer

The instrument is used for disinfectant measurement in drinking water, swimming pool and industrial cooling water, water plant effluent and municipal pipe network monitoring.

Parameters Of Chlorine Dioxide Analyzer

Product Name Chlorine dioxide analyzer
Product Model

ADS2-U3A-CD7

Software Version

DRFN chlorine dioxide Analysis software V1.0

Display Range

0.00-200.00mg/L

Resolution

0.001 mg/L

Accuracy

Better than 0.15% of the measuring range

Stability

0.05% of the measuring range every 24 hours, no accumulation

Repeatability

Better than 0.1% of the measuring range

Relay Control

2 settable SPST relays, the max. load 3A/250VAC

Sensor Self-diagnosis

screen image and buzzer prompt

Analog Current Output

2 settable 0/4~20mA active current loop output max. load 1000Ω

Communication Method

RS485, MODBUS RTU; support JSON text data format

Display Screen

128*64 LCD

Time History/Data Record

Year/month/day/hour/min/second, record 14000 historical data

Protection Grade

IP66

Operating Environment

0~60℃, relative humidity less than 95%

Storage Environment

-20~70℃, relative humidity less than 55%

Shell Material

Enhanced ABS

Dimensions

144*144*120mm (hole size 138*138 mm)

Mounting Method

Wall mounting, pipe clamping, panel installation

Power Supply

100~240V AC

Weight

800g

Power Consumption

12W

11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane

The Matching Chlorine Dioxide Sensors Recommendation

  • Model: CD7
  • pH range: 1~11
  • Pressure range: 0~1 bar, no pressure surges and/or fluctuations
  • Temperature range: 0~ 50℃

  • Integrated automatic temperature compensation

  • Response time: T90 about 90s
  • Flow rate: about 30 l/h, low flow-dependence
  • Shaft length: standard 175 mm, and up to 220 mm in length (mA-Version)
  • Connection: for mA-version 2-pole terminal, M12 male

  • Material: PVC-U, stainless steel 1.4571, semipermeable membrane

Item number Type Measuring range:
ppm
Resolution:
ppm
Output signal Power supply
3226620 CD7H-M0c 0.005~2.000 0.001 Modbus RTU

9~30 VD

20~56 mA

3226621 CD7N-M0c 0.05~20.00 0.01
3226622 CD7L-M0c 0.5~ 200.0 0.1

Send your message to this supplier
Send Now

11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane

Ask Latest Price
Brand Name :
Daruifuno
Model Number :
ADS2-U3A-CD7
Certification :
CE
Place of Origin :
China
MOQ :
Negotiable
Price :
Negotiable
Contact Supplier
11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane
11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane
11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane
11pH Ampere Chlorine Dioxide Analyzer With Surfactant Resistant Membrane

Suzhou Delfino Environmental Technology Co., Ltd.

Verified Supplier
5 Years
jiangsu, suzhou
Since 2015
Business Type :
Manufacturer, Exporter, Seller
Certification Level :
Verified Supplier
Contact Supplier
Submit Requirement