• Flammable substances (such as alcohol based skin prepping agents and tinctures)
• Naturally occurring flammable gases which may accumulate in body cavities such as the bowel
• Oxygen enriched atmospheres
• Oxidizing agents (such as nitrous oxide [N2O] atmospheres)
The sparking and heating associated with electric surgery can provide an ignition source. Observe fire precautions at all times. When using electric surgery in the same room with any of these substances or gases, prevent their accumulation or pooling under surgical drapes, or within the area where electric surgery is performed.
This unit is a plasma electrosurgical system with bipolar and multi-polar functions. Mainly containing the parts as follows:
1). One tabletop generator device.
2). One reusable and non-sterile power supply cable.
3). One reusable and non-sterile footswitch.
4). Disposable Plasma probes.
PLA-600 Generator Output Characteristics
Max. (open circuit) Peak voltage | Max. VRMS Voltage | Rated Load (ohms) | Max. Output Power (W) | |
ABLATION1 | 680 | 295 | 250 | 348 |
ABLATION2 | 620 | 274 | 250 | 300 |
PLACOAG | 380 | 158 | 250 | 100 |