Orange Color Plasma Surgery System Ablation for Urology Surgery
Warnings
- Do not operation as following instructions will result in serious surgical consequences.
- Do not use non-conductive liquid (such as: distilled water, glucose, air, gas, glycine, etc.) as a medium, only conductive liquid available, such as saline or lactic acid Ringer liquid.
- Do not place working parts near or near combustible materials such as gauze or surgical shades.
- Activated or in hot condition general surgery parts may causing fire.
- Probe tip temperature may hot enough to cause burns after cut off of surgery current.
- Move or activate the probe casually out of sight may cause injury to patients.
- The current caused by other equipment or conductive objects may cause local burns to the patient or physician.
- Direct touch of working electrode, either put the working electrode or loop electrode closing to conductive object may conduct current flow through the object.
- If the overheat or excessive external force damages the tip, it can cause alien fragmentation and may require extensive surgical resection.
- Electrical shock: Do not connect wet attachments to the main unit.
After Surgery
- Do not use flammable materials to clean and disinfect host or cable.
- Do not open the cover of the machine. Find qualified personnel to repair.
- Replace the main unit's fuse with only the same type and grade of fuse.
- A host failure may be caused by unintentional increase in output power.
- The therapy device is designed to operate independently as a unit and can only be used by the same company's accessories.
PLA-600 Generator Output Characteristics
| Max. (open circuit) Peak voltage | Max. VRMS Voltage | Rated Load (ohms) | Max. Output Power (W) |
ABLATION1 | 680 | 295 | 250 | 348 |
ABLATION2 | 620 | 274 | 250 | 300 |
PLACOAG | 380 | 158 | 250 | 100 |