Snoring Treatment Plasma Surgery System Minimally Invasive Ablation
Trouble-shooting
The PLA-700 includes automatic self-diagnostics. If the diagnostics detect an error, the system sounds an audible tone, and deactivates the unit output power.
Most errors result from faults in accessories (like treatment handle/probe, footswitch) attached to the unit. The following table lists the issues, describes the errors, and recommends actions to take to resolve the errors.
If the unit displays any other error, it requires service.
Fuse Replacement
Fuses for the unit reside directly below the Power Cable Receptacle on the rear of the unit.
To replace the fuses, follow this procedure:
- Unplug the power cord from the wall outlet.
- Remove the power cord form the Power Cable Receptacle on the rear panel.
- To release the fuse drawer, insert a small flathead screwdriver into the slot on the drawer below the power cord receptacle. Then, slide the drawer out.
- Remove the two fuses and replace them with new fuses with the same values/specifications refer to section 7 system specification.
- Insert the fuse holder into the Power Cable Receptacle.
Periodic inspection
Every six months, visually inspect the PLA-700 for signs of wear or damage.
In particular, look for any of the following problems:
- Damage to the power cord
- Damage to the power cable receptacle
- Obvious damage to the unit
- Damage to any receptacle
- Accumulation of lint or debris in or around the unit.