Three Phase Energy Meter Testing Equipment high accuracy 0.01 IEC standard
KE8300F Three Phase Energy Meter Testing Equipment
Accuracy : 0.01 , Used for the test of standard meter and mounted meter with a accuracy class below 0.05, Number of meter position : 3 - 12
Features:
· High accuracy: 0.01%
· Integrated / separated aluminum alloy structure, light and reliable, flexible choices
· Meter positions can be chosen from 3 - 12
· Auto protection of short circuit of voltage, open circuit of current, and overload.
· Full automatic, semi automatic and manual operation available
· Supporting bar code input, which increase the efficiency of input
· Simultaneous test of the meters with same specification, different constants and different classes.
· Forward and reverse active energy / forward and reverse reactive energy error test
· Creeping and starting test
· Standard deviation, 24 hour variation test
· Timing error test
· Influencing quantity test of voltage, frequency, harmonic
· Calibration and verification of constant
· Register indicating value, register combination error test
· Powerful computer management system, easy and user-friendly
· Perfect database management makes the storage, searching, and viewing of the data possible, and the data can be converted and exported in other forms
Main Specification:
KE8300F |
|
Number of meter position |
3/6/10 |
Accuracy |
±0.01% |
Voltage output |
|
Output range (P-N) |
3 x (24V - 300V) |
Output power of voltage |
3 x 200VA |
Resolution |
Better than ± 0.01% |
Stability |
Better than ± 0.02% / 150s |
Load regulation rate |
Better than ± 0.01% |
Distortion factor |
Better than ± 0.3% (linear resistance load) |
Harmonics output |
2nd -21st free programmable |
Current output |
|
Output range |
3 x (1mA - 120A) |
Output power of current |
3 x 300VA |
Resolution |
Better than ± 0.01% |
Stability |
Better than ± 0.02% / 150s |
Load regulation rate |
Better than ± 0.01% |
Distortion factor |
Better than ± 0.3% (linear resistance load) |
Harmonics output |
2nd -21st free programmable |
Phase output |
|
Output range |
0 -- 360° |
Resolution |
± 0.01° |
Frequency |
|
Output range |
45Hz - 65Hz |
Resolution |
± 0.01Hz |
Operation condition |
|
Power supply |
220V±10%,50/60Hz±2Hz |
Temperature |
-10°C -- +40°C |
Humidity |
35% -- 85%RH |