1ph Electric Meter Test Equipment multiple current channels class 0.05 120A
KE8102 single phase meter test bench
KE8102 has multiple switchable channels for current output, used for testing single phase mechanical and electric meter that have 2 or 3 elements, Number of meter position : 3-64
Features:
· Integrated / Separated aluminum alloy structure, lightand reliable, flexible choices
· The numberof meter positions can be chosen from 3 - 64
· Perfect alarming protective function, auto protection ofthe short circuit of voltage, the opencircuit of current, and overload.
· Full automatic, semi automatic and manual operationavailable
· Supporting bar code input, which increase the efficiencyof input
· Simultaneous test of the energy meters with same specification, different constant and different class.
· Auto mark locking increases the efficiency of test
· Forward and reverse active energy / forward and reversereactive energy error test
· Creeping and starting test
· Standard deviation, 24 hour variation test
· Influencing quantity test of voltage, frequency, harmonic
· Calibration and verification of energy constant
· Powerful computer management system, easy anduser-friendly
· Perfect database management makes the storage, searching,and viewing of the data possible, and the data can be converted and exported inother forms
· The testof open circuit for Neutral voltage (optional)
· Adopt multi-serial port system, realizing the parallelcommunication of computer andthe tested meters (optional)
Main Specification:
KE8102-48 | |
Number of meter position | 48 |
Accuracy | ±0.05% |
Voltage output | |
Output range (P-N) | 24V - 300V |
Output power of voltage | 1000VA |
Resolution | Better than ± 0.01% |
Stability | Better than ± 0.02% / 150s |
Load regulation rate | Better than ± 0.01% |
Distortion factor | Better than ± 0.3% (linear resistance load) |
Harmonics output | 2nd -21st free programmable |
Current output | |
Output range | 1mA - 120A |
Output power of current | 2000VA |
Resolution | Better than ± 0.01% |
Stability | Better than ± 0.02% / 150s |
Load regulation rate | Better than ± 0.01% |
Distortion factor | Better than ± 0.3% (linear resistance load) |
Harmonics output | 2nd -21st free programmable |
Phase output | |
Output range | 0 -- 360° |
Resolution | ± 0.01° |
Frequency | |
Output range | 45Hz - 65Hz |
Resolution | ± 0.01Hz |
Operation condition | |
Power supply | 220V±10%,50/60Hz±2Hz |
Temperature | -10°C -- +40°C |
Humidity | 35% -- 85%RH |
Production area