OEM 3/5/12 Lead Temperature Respiration ECG Module for Patient Monitor
Adopting digital filter technology |
Selectable 50Hz or 60Hz band stop filter |
500Hz,7 /12 lead synchronous sample and analysis |
Character Classification Up to twelve sort, and 3-channel ST Segment Monitoring |
Anti high-frequency electric-knife |
Adopting single 5V power supply,low consumption and safely insulation |
5V TTL Communication |
2-channel temperature detect ,and two sort of sensor model |
Detecting respiration by impedance between RA-LL |
8-channel ECG waveform |
ST Segment Analysis |
Heat rate calculation |
Arrhythmia analysis |
Lead off status indication |
Lead mode | 12 leads (R, L, F, N, C1~C6 or RA, LA, LL, RL, V1~V6 ) |
Selection | I, II, III, aVR, aVL, aVF, Vx, CAL |
Gain | ´0.25V/mV, ´0.5V/mV, ´1V/mV, ´2V/mV |
Heart rate(HR) | Range: Adult 15 ~ 300 bpm Neo/Ped 15 ~ 350 bpm Accuracy: ±1% or ±1bpm Sensitivity: > 200 uVpp |
Differential Input Impedance | > 5 MΩ |
CMRR | > 100 dB |
Electrode offset potential | ±400mV |
Insulation Intensity | 4000VACrms |
Baseline Recovery | < 3 s, after Defi. |
ECG Signal Range | >±5 mVpp |
Bandwidth | Diagnostic 0.05 ~ 100 Hz Monitor 0.5 ~ 40 Hz Surgery 1 ~ 20 Hz Strong filter 5 ~ 20Hz |
Calibration Signal | 1mVpp,Accuracy ±3% |
ST segment monitoring range | -2.0mV ~2.0mV |
Arrhythmia analysis | ASY,VF/VTA,CPT,RUN,BGM,TGM,R on T(ROT),VPB,TAC,BRD,MIS,PNC,PNP |
Waveform | 1 ch |
Respiration Rate | dult 7-120bpm Ped and Neo 7-150bpm Resolution 1bpm Accuracy ±2bpm |
Suffocation alarm | 10-40s Gain 10levels |
Baseline Impedance | 500 - 4000Ω |
Measuring Range | 0.5 - 5Ω |
Channel Number | 2 ch |
Probe type | 2K and 10K probe |
Range | 0-50°C |
Resolution | 0.1°C |
Accuracy | ±0.1°C |