Vertical laminar flow clean bench is a laminar flow work cabinet or similar enclosure that provides filtered air across the work surface to protect against contamination.
The vertical laminar flow clean bench was originally created to supplement clean room technology and today is used across a broad spectrum of industries including research, manufacturing, aerospace, bioscience, pharmaceutical production and food processing.
Many medical preparations take place on a clean bench, which explains why you often see them hospitals, clinics, laboratories and medical facilities worldwide.
Simply put, all air within the table envelope flows in a single direction with no backwash, which is usually an area of turbulent airflow created by obstacles, equipment or movement within the cleaning table enclosure. Proper housing design helps to prevent backwashing and ensures that vertical laminar flow continues to protect the work of the cleaning bench surface.
Features
1. Materials choice of color plate steel,
2. Consists of a certified fan filter module,
3. HEPA Filter rated 99.99% efficient,
4. Built-in fluorescent light.
Parameters
Item | GCC-VLBC-960 | GCC-VLBC-1200 | GCC-VLBC-1500 |
Internal Size | L930*W600*H600mm | L1170*W600*H600mm | L1470*W600*H600mm |
External Size | L960*W780*H1580mm | L1200*W820*H1580mm | L1500*W900*H1580mm |
Cleanliness Level | Level 100 | ||
Wind Speed | 0.3~0.6m/s | ||
Noise | <65dB | ||
Air Flow | 1200m³/h | 1800m³/h | 2300m³/h |
Efficiency | 99.99%@0.3μm | ||
Voltage | 220V/50Hz | ||
Power | 280W | 405W | 490W |
Options
1. Stainless steel attached worksurface,
2. Stainless steel cleanroom table,
3. Stainless steel interior side panels,
4 .UV sterilization light and ionizing bar.
Suggestion
1. Pre-filter: each one should be replacement in each 6 month, but it can refresh no more three times.
2. HEPA filter: each one should be replacement in each half and one years.