1. Product Information
Insulating oil tester calibrator system JY-100
2. Equipment Description:
1 Breakdown voltage error calibration for insulating oil tester
2 High voltage measurement error calibration for insulating oil tester
3 Voltage rise rate error calibration for insulating oil tester
4 High voltage cut off time measurement for insulating oil tester
5 High voltage distortion rate measurement
3.Technical Parameters:
1 High voltage measurement: 0-80kv/0-100kv
2 High voltage measurement error: 0.2%RDG+0.02kv
3 High voltage rise rate measurement error < 0.5%RDG+0.005kv/s
4 High voltage distortion rate measurement error <2%
5 Cut off time measurement error <0.08ms
6 Measurement high voltage divider capacity <20pf
7 Power supply: AC220V±10% 50Hz/60Hz Weight < 10kg
8 Environment temperature: 0~50°C Humidity: ≤80%RH
4. Product Picture: