100L Visual Leak Detector 8m Negative Pressure Test Machine Real-time Observation
Product size and material:
The detail images:
exhaust valve
the edge of lid
Test principle:
IPX8 test:
Positive pressure leak test:
Negative pressure leak test:
Product outline:
This is a visual leak detection machine equipped with a negative pressure function, with a maximum negative pressure of up to 8 meters, suitable for leak detection in pipelines, containers, and other equipment under various conditions. The device features a simple design and supports manual lid opening, making it convenient for users to operate and maintain before and after testing.
Standard features:
Item | Parameters |
Thickness of lid | 25mm |
Flange thickness | 30mm |
Tank thickness | 15mm |
Equipment net weight | Approx 120kg |
Pressure range | -50kPa ~ 300kPa |
Water depth range (negative pressure) | 8 meters |
Adjustment method | Automatic adjustment |
Pressure accuracy | 1 kPa |
Pressure error value | ±2kPa |
Set pressure of safety valve | 0.35Mpa(350kPa) |
Testing time9 | 0-999999S |
Power supply | 220V/50Hz |
Power | 100w |
Working scene:
Visual leak detectors are integral to accurately identifying leaks in systems like pipelines, pressure vessels, and heat exchangers, particularly in applications demanding precision. Their use in industrial production, maintenance, and quality assurance ensures greater efficiency and reliability in the leak detection process.
Remark:
1 The pictures in this document are for customer reference only, and there may be color differences between the actual model and the pictures.
2 This document only lists some test standards, which does not mean that other test requirements and standards cannot be conducted. Welcome to communicate and discuss with us.
3 One test only at the same time. The next project can be only carried out after the current test is completed. In addition, we can provide multi-functional customized test solutions according to user requirements.
4 The formula of the corresponding relationship between pressure and water depth: P=9.8d (Unit :kPa,m).
5 Sometimes there are gaps on the surface of the product's shell, but this part will not last for a long time or bubbles occasionally. Usually, the condition for us to judge whether it is leaking is: observe continuous bubbles.If the product is small and there is almost no space inside (such as lamp beads), the judgment standard is stricter:: that is, if there are air bubbles, it will be judged as leakage.
6 Negative pressure testing is suitable for testing samples whose internal pressure is higher than the external ambient pressure, or for quick testing (no inflation process).
7 It is necessary to avoid scratching the inner wall of the container by the sample, as scratches will affect the observation and video effects.
8 At present, only Chinese and English interfaces are supported; if other languages are introduced, the software can be updated remotely.
9 The time unit set by the factory is generally in seconds. If you want to extend the test time, you need to explain in advance.
10 The photos may not match the actual ones, and are for user reference only.
About us:
Tags: Leak Visualization Machine , Leak Inspection Machine with Visual Display , Leakage Detection Device