Electrical Conductivity Meter, Eddy current electrical
conductivity Tester and meters REC-102

REC-100 series of eddy current electrical conductivity meters, which are designed for quick &
convenient measurement of material property, such as material separate,quality control,material
state check and so on. The products use the electromagnetism testing principle, can testing objects
focus on non-ferromagnetic materials.
REC-100 series products use 60 KHz (aviation industry standard) to inspirit, the testing data can be
read in two kinds of unit (%IACS and MS/m). The products can test electrical conductivity and resistivity
at the same time, and they include the intelligent temperature and lift-off compensations, thus can
guarantee the testing accuracy of each measurement. The big typeface and the back light illumination
designs are advantageous for users to take the testing data even in low light condition. The products
use the high property battery to make sure it keeps more running time, and because of their mini-size,
you could easy to carry and to grasp hold. The design of the products are more advantages, such as
users can replace the probe in the outdoor, don’t need to return the company to adjust the probe into
matching the meter.
It can store 1000 measurement data, moreover, users can use USB connect
the meter to a PC to create testing report.
Application:
★ Absolute measurement of electrical conductivity of nearly all non-ferrous metals;(Ti – Ag)
★ In the aerospace and automotive industry,monitor the process of heat treatment, the strength and
hardness of aluminum alloy;
★Test the conductivity when the aluminum have not been oxidized;
★ Measurement the purity of non-ferrous metals;
★Test the resistivity of conductive materials;
★Analyze thermal properties of materials.
REC-102 Technical parameters
NAME | CONTENT | NOTE |
Measurement technology | Eddy current | |
Operating frequency | 60KHz | |
Display screen | 240X320 pixels TFT-LCD | 4 kinds of background color |
L*B*H | 180*80*35 mm | |
Instrument case | Anti-intense impact,water-proofing polyester | Match the fixed belt the leather protective covering |
Weight | 245g | Contain battery |
Power supply | High capacity, high performance lithium polymer battery | Not back light illuminate may keep on working about 100 hours. |
Measuring range | 1%IACS—105%IACS(0.58 MS/m -61MS/m) | |
Distinguishing rate | 0.01% IACS | |
Probe compensated | 0.05mm | |
Measuring accuracy | 0°C to 50°C 23%IACS :±0.25%IACS 100%IACS :±0.5%IACS | |
Temperature compensation | 0°C~50°C | Automatic compensation function |
Normal work environment | Relative humidity | 0~95% | |
Operating temperature | 0℃~50℃ | |
Language | Chinese, English | |
PC communication | USB | |
Fitting | Portable box; probe; probe in cable; operating manual; conductivity standard sample; adapter | Conductivity sample installs in the instrument, to verify and guarantee temperature compensation |
Probe Diameter | 12.7mm (Applicable to minimum measuring area diameter at 60KHz is13mm.) | |
Reading storage | Preserve 1000 survey data | |
Hand-held eddy current conductivity tester for NF-metals with automatic normalization of the conductivity values to 20°C through temperature sensor.
TAG: Eddy Current Conductivity Tester