The 5560A High-Performance Multi-Product Calibrator calibrates the most popular benchtop 6.5-digit multimeters with better than 4:1 test uncertainty ratio (TUR), oscilloscopes to 2 GHz, and clamp meters to 1500 A.
Technical improvements include 30 A continuous current output with no duty cycle and synthesized inductance functionality. These improvements enable you to calibrate modern and full-functional benchtop multimeters and clamp meters. A large variety of thermocouple types and a new thermocouple connector enables easier connection and broadens the temperature workload coverage. The workload is expanded even further when the 5560A is used with a 52120A amplifier and appropriate coil to increase the current output to 6000 A, for calibrating high-current devices such as Rogowski coils.
Highlights of Fluke 5560A Multi Product Calibrator
5560A: The next generation high-performing calibrator supporting the widest workload coverage
The 5560A High-Performance Multi-Product Calibrator defines a new class of high-performance multi-product calibrator, providing the broadest electrical workload coverage and highest accuracy for calibrating the most demanding modern workloads.
Key features and benefits of 5560A Calibrator
Function | Range |
Direct volts | 0 V to ± 1020 V |
Direct current | 0 to ± 30.2 A |
Alternating volts | 1 mV to 1020 V 3 Hz to 500 kHz |
Alternating current | 10 μA to 30.2 A 3 Hz to 30 kHz |
Volt/hertz | 1000 V at 10 kHz/330 V at 100 kHz |
Waveforms | Sine wave, triangle wave and truncated sine wave |
Resistance variable | 0 Ω to 1200 MΩ |
Capacitance variable | 200 pF to 120 mF |
Inductance variable | 12 uH to 12 H |
Power (dual output – phantom loads) | 30.9 kW |
Oscilloscope bandwidth | 2.1 GHz, 1.1 GHz, or 600 MHz |
Phase lock | Yes |
Phase control | 0.01° |
Frequency uncertainty | <2.5 ppm |
Temperature standard | <ITS-90, IPTS-68 |
Resistance temperature detectors (RTDs) output | Pt100 385, Pt100 395, Pt100 3916, Pt200 385, Pt500 385, Pt1000 385, Ni120, Cu10 Compensation: off, 2-wire, 4-wire |
External frequency reference 10 MHz | Yes |
Thermocouples (source and measure) External or internal CJR | B,C,E,J,K,L,N,R,S,U, BP and XK 10 μV/°C |
Scope calibrator | SC600 – with levelled sine up to 600 MHz SC1100 – with levelled sine up to 1100 MHz SC2100 – with levelled sine up to 2100 MHz |
Interfaces / Remote commands | USB 2.0, Etherne Telnet, RS-232 and GPIB |
52120A Amplifier current capability | Current capability up to 120 A Current capability up to 360 A with three 52120As (parallel) |
List of Options of FLUKE 5560A
5560A/600M | 600 MHz Oscilloscope Calibration Option |
5560A/1G | 1.1GHz Oscilloscope Calibration Option |
5560A/2G | 2.1GHz Oscilloscope Calibration Option |
FAQ:
1.Can you ship order to Canada or Australia ?
Yes,we can .We send orders worldwide .