Laboratory Computer Servo Rubber Rotorless Moving Die Rheometer
Use:
This machine is designed by GB/T16584<Rubber——use no-rotor rheometer to test vulcanization characteristic,can meet ISO6502 standard,and can test the data of Italy standards about T10,T30,T50,T60,T90.This machine can be use test unvulcanized rubber characteristic,and test out the suitable vulcanizing time. Used imported intelligent digital temperature controller,easy adjustment,wide temperature controlling range,high accuracy,good stability and reproducibility.Use computer to control the machine and store the test data and print the data.Test curve can be compared and amplification in software.
Main functions:
non-curing equipment used monolithic rotor control, which include: host, temperature measurement, temperature control, data acquisition and processing, sensors and electrical chains and other components. These measurements, temperature control circuit consists of a temperature control device, platinum resistance, heater composition, capable of automatic tracking power and ambient temperature changes, automatically correct PID parameters to achieve fast and accurate temperature control purposes. Data acquisition system and mechanical linkage to complete the rubber vulcanization process of force torch signal automatic detection, automatic real-time display of temperature and settings. After curing, automatic processing, automatic calculation, print vulcanization curve and process parameters. Show curing time, curing power Ju, also has a variety of audible alert.
non-rotor curing equipment controlled by computer, the computer setting the parameters of the direct control of test parameters rheometer. Display real-vulcanization curve and temperature curve, store test results, different adjustable comparison of test results and in a different color.
Features:
1. This instrument is really confined mold cavity, and the United States Alpha (formerly Monsanto) has done. Repeatability, and test data comparable with the Alpha. At the international leading position in the same industry.
2. The instrument development platform based on large databases, temperature control devices using the software to directly control and collection and processing. Instrument to overcome the general curing temperature of the shortcomings of using temperature controller (accuracy poor). The technical leadership of international new trend.
3. This instrument has statistics, analysis, storage and comparison functions. Humanized design, easy to operate
4. Using imported high-precision sensors
Download Brief:
ZL-3001 No Rotor Rheometer.pdf
The main technical parameters:
Model | Without Rotor Rheometer Rubber Testing Machine |
Temperature tester | Room temperature to 200°C |
Temperature accuracy | ≤ ±0.3°C |
Temperature resolution | 0.1°C |
Torque range | 0-5N.M,0-10N.M,0-20N.M |
Torque resolution: | 0.001NM |
Motor rotating speed | 100 times / min |
Rotor range | Adopt 4 setting: 25, 50, 100, 200lbs per inch |
Die cavity frequency | 1.7Hz |
Sway angle | ±0.5°, ±1°, ±3° |
Air compression | 0.5Mpa--0.65MPa |
Ambient temperature | 0-35°C |
Test result | Printing:date,time,temperature,vulcanization curve,temperature curve,ML,MH,ts1,ts2,t10,t50,Vc1,Vc2. |
Dimension | 131*56*62cm |
Power | Single phase, 220V±10%, 50Hz |
Display | Windows 7 + software |
Weight | 250kg |
Using computer control, after setting the parameters on the computer, directly control the test parameters of the vulcanizer. Real-time display of vulcanization curve and temperature curve, storage of test results, can adjust the comparison of different test results and display them in different colors.