A robust and portable single channel meter with light duty Dissolved Oxygen probe kit designed for field use with simple USB data transfer capabilities
The HQ1130 Portable Dissolved Oxygen Meter is designed for water quality professionals who need to perform dissolved oxygen analysis in either field and lab environments. This new portable DO meter and probe kit allows you to collect accurate measurements, manage data, and easily review results, and with its IP67-rated ruggedness, it can withstand field use.
Specifications
Accessories Included | USB charger, power cable, and clips |
Automatic Buffer Recognition | Yes |
Backlight | Yes |
Barometric Pressure Measurement | Automatic compensation of DO when using an LDO or LBOD probe |
Cable Length | 1 m |
Calibration Intervals/Alerts/Reminder | Off, selectable from 2 hours to 7 days |
Certifications | CE, FCC, ISED, RCM, KC, ETL Verified: US DOE/ NRCan Energy Efficiency, RoHS |
Data Export | USB connection to PC or USB storage device (limited to the storage device capacity). |
Data Memory | 5,000 data points |
Data Storage | Automatic in Press to Read Mode and Interval Mode. Manual in Continuous Read Mode. |
Dimensions (H x W x D) | 63 mm x 97 mm x 220 mm |
Display | Up to 3 parameters at a time, dependent on HQ model |
Display Type | 536x336 Mono-TFT |
DO Measurement Range | 0.1 - 20.0 mg/L (ppm), 1 - 200% saturation |
DO Resolution | 0.1 |
Enclosure Rating | IP67 (with battery compartment installed) |
GLP Features | Date; Time; Sample ID; Operator ID, Calibration |
Inputs | 1 |
Instrument | Portable |
ISE Direct Measurement Range | No |
ISE Electrode Calibration | No |
Kit? | Yes |
Languages | English, Chinese - Simplified, Spanish, French, German, Italian, Japanese, Turkish, Dutch, Portuguese (PT & BR), Korean, Danish, Swedish, Polish, Norwegian, Hungarian, Greek, Finnish, Czech, Romanian, Croatian, Bulgarian, Slovak, Lithuanian, Estonian, Slovenian, Russian |
Lock Function | Continuous / Auto-stabilization ("press to read") / At Interval |
Measurement Method | Probe specific programmed method settings |
Model | HQ1130 DO/1 Channel |
mV Measurement at Stable Reading | No |
mV Measurement Range | No |
mV Resolution | 0.1 mV |
Operating Error Messages | Clear text error messages displayed |
Operating Humidity | 90% (non-condensing) |
Operating Interface | Soft Touch Keypad |
Operating Temperature | 0 - 60 °C |
Parameter | Dissolved Oxygen (DO) |
Power Supply | Rechargeable lithium-ion battery 18650 (internal) |
Class II, USB power adapter: 100 - 240 VAC, 50/60 Hz input; 5 VDC at 2 A USB power adapter output (external) | |
Probe Type | Intellical Standard Laboratory or Rugged Field |
Probes included | LDO10101 |
Protection Class | IEC Class III (SELV (Separated/Safety Extra-Low Voltage) powered); external |
Sensor A | LDO10101 |
Sensor B | None |
Sensor C | None |
Storage Conditions | -20 - 60 °C, max. 90% relative humidity (non-condensing) |
Temperature Compensation | Automatic Temperature compensation for pH |
Temperature Measurement | °C or °F |
Temperature Resolution | 0.1 °C |
Type of measurement | Field Measurements |
Warranty | 12 months |
Weight | 519 g (Meter only) |
What's included? | LEV015.53.1130A HQ1130 Portable DO Meter, 1 channel |
LDO10101 IntelliCAL dissolved oxygen (DO) sensor, 1 m cable | |
LEZ015.99.A001A HQ standard electrode field case | |
LEZ015.99A003A HQ hand strap & kickstand | |
LEZ015.99.A004A HQ protective glove | |
User manuals |