Model | T2600 | K8000 | |||||||||
Optical System | Double Beam, Grating 1200 lines/mm | ||||||||||
Wavelength Range | 190-1100 nm | ||||||||||
Spectral Bandwidth | 2 nm | ||||||||||
Wavelength Accuracy | ±0.1nm@656.1nm, ±0.3nm@all | ||||||||||
Wavelength Repeatability | ≤0.1 nm | ||||||||||
Photometric Accuracy | 0.2%T(0~100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A) | ||||||||||
Photometric Repeatability | ≤0.15%T (0-100%T), 0.001A(0-0.5A), 0.002A(0.5-1A) | ||||||||||
Photometric Range | 0-200%T, -0.3~3A, 0-9999C (0-9999F) | ||||||||||
Stray light | ≤0.03%T@220nm,360nm | ||||||||||
Stability | ±0.0005A/h@500nm | ||||||||||
Baseline Flatness | ±0.002A | ||||||||||
Noise | 0.0005A@500nm | ||||||||||
Work Mode | T, A, C, E | ||||||||||
Scanning Speed | Hi, Med, Low (Max. 3000nm/min) | ||||||||||
Wavelength Setting | Auto | ||||||||||
Display | 7" TFT Color Screen | ||||||||||
Light Source | Imported Deuterium & Tungsten lamp | ||||||||||
Detector | Imported Silicon Photodiode | ||||||||||
Cuvette Holder | 10mm manual 4-cell holder | ||||||||||
Output | USB drive, USB host, RS232 | ||||||||||
Power | AC 220V/50Hz; AC 110V/60Hz | ||||||||||
Shipping Size | 810*660*390mm | ||||||||||
Gross Weight | 27 kg | ||||||||||
Standard Accessaries | 10mm glass cuvette x 4,10mm quartz cuvette x 2, Power cord, User's Manual,PC software |