IEC 60335 Dual Station Label Marking Abrasion Test Apparatus
Standards For Dual Station Label Marking Abrasion Test Apparatus:
This label marking abrasion test apparatus conforms to IEC 60335-1 clause 7.14, IEC 60065:2014 clause 5.1, IEC 60950:2013 clause 1.7.11, IEC 60669-1 clause 8.9.
Technical parameters For Dual Station Label Marking Abrasion Test Apparatus:
Test stations | Two stations |
Test stroke | 10-60mm, adjustable(Manually adjustable) |
Speed | 5-60 times /min, adjustable |
Counter | 0-9999 adjustable |
Weight | Two pieces of 10g, 20g,50g, 100g, 200g, 500g |
Voltage | AC 220V, 50Hz |
Machine size: | 500*300*550mm |
Maintenance Notes For Dual Station Label Marking Abrasion Test Apparatus:
1 The instrument must be placed on a stable surface before use.
2 It is not allowed to move the machine during use.
3 Select the appropriate power supply voltage, should not be too high, to avoid burning the instrument.
4 If the instrument works abnormal, please contact with us for timely processing.
5 The instrument must be used in a good ventilation environment.
6 Please add lubricating oil to the mechanical transmission part in time due to the large operating load of each component of the button mechanism.
7 After each test is completed, operator should clean the machine.
8 The control box part must be wiped with a dry cloth.
Troubleshooting For Dual Station Label Marking Abrasion Test Apparatus:
1 If the POWER is turned on, the instrument is not powered on or the power indicator is not light: if the power indicator is damaged, replace the indicator; if there is no power output at the power output terminal, please check the power cord or power socket-outlet.
2 If the power is turned on, the power indicator lights, but the machine does not work when operator presses the SWITCH button. Please check whether there is a disconnection. If there is a disconnection, please connect it in time. If there is no disconnection of the relay, indicating that the counter is damaged, the counter should be replaced in time.
3 When the number of times set on the counter is finished, but the motor does not stop working, indicating that the counter is damaged.
4 If the power indicator light is on but the counter has no display, it means that the counter is damaged and the counter needs to be replaced.
Configuration
No.: | Name | Unit | Quantity | |
1 | Main device | Label Marking Abrasion Test Apparatus | Set | 1 |
2 | Accessory | Petroleum spirit/ alcohol abrasion test fixtures | PCS | 2 |
Rubber head abrasion test fixtures | PCS | 2 | ||
Pencil abrasion test fixtures | PCS | 2 | ||
Cotton cloth | Pack | 1 | ||
Rubber head | PCS | 2 | ||
Pencil | PCS | 2 | ||
3 | Document | User manual | Copy | 1 |
Certificate of quality | Copy | 1 | ||
Warranty card | Copy | 1 |