Specifications
Brand Name :
Sinuo
Model Number :
SN2801
Certification :
Calibration Certificate (Cost Additional)
Place of Origin :
China
MOQ :
1
Price :
Negotiable
Payment Terms :
T/T, L/C, Western Union, MoneyGram
Supply Ability :
10 Sets Per Month
Delivery Time :
20 Days
Packaging Details :
Plywood
Applied standards :
IEC60601-1:2012
Specimen and Limitations :
Moving ME equipment or moving parts
Station :
Single station, can be customized
Test size range :
W80cm; H100cm
Drive mode :
Servo motor + precision reducer drive
Linear stroke :
0-3600mm, can be preset
Test speed :
0-0.85m/s can be preset, can be preset (0.8±0.1m/s)
Test count :
0~999999 times, can be preset
Door frame :
WDH=40mm×40mm×1500mm
Safety protection :
Fences all around
Description

IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus

Product overview:

This Impact Testing Apparatus is used for moving medical electrical equipment or moving parts of medical electrical equipment.

It can evaluate whether the mechanical strength of medical electrical equipment, or moving parts of medical electrical equipment, is subject to unacceptable risk to the equipment when subjected to rough handling and movement. The motor-driven manipulator pushes the sample at a speed of 0.8m/s±0.1m/s to realize the upper step impact test, the lower step impact test and the door frame impact test on the water platform.

The device includes an elongated work table (track), with a 40mm high step in the middle of the table, and a 40mm wide hardwood at the end; including a servo motor, a reducer and a guide rail drive mechanism. Equipment dimensions: about: 4000 × 1300 × 1850 (mm) W × D × H, equipment weight: 350kg

Technical parameters:

Station Single station, can be customized
Electrical control method PLC intelligent integrated control, color man-machine touch interface operation
Test size range

The distance between casters is not more than 80cm, the width is not more than 80cm, and the height is not more than 100cm

Drive mode Servo motor + precision reducer drive
Specimen clamping method Mechanical clamping
Test items and action process

1. Impact test on steps: move from left to right, impact the middle obstacle

2. Down-step impact test: move from right to left and push down from the middle obstacle

3. Door frame impact test: move from the middle to the right, impact the door frame obstacle

Linear stroke 0-3600mm, can be preset
Test speed 0-0.85m/s can be preset, can be preset (0.8±0.1m/s)
Test count 0~999999 times, can be preset
Upper step impact test obstacle 40mm steps, located in the middle of the platform, width 1000m
Descending Step Impact Test Obstacles
Door frame impact test obstacle WDH=40mm×40mm×1500mm, located at the right end
Safety protection Fences all around

Send your message to this supplier
Send Now

IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus

Ask Latest Price
Brand Name :
Sinuo
Model Number :
SN2801
Certification :
Calibration Certificate (Cost Additional)
Place of Origin :
China
MOQ :
1
Price :
Negotiable
Contact Supplier
IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus
IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus
IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus
IEC60601-1 Medical Electrical Equipment Impact Testing Apparatus

Sinuo Testing Equipment Co. , Limited

Verified Supplier
7 Years
guangdong, guangzhou
Business Type :
Manufacturer
Total Annual :
800000-1500000
Employee Number :
30~50
Certification Level :
Verified Supplier
Contact Supplier
Submit Requirement