Pit Mounted For Direct Floor Loading Horizontal Pharmaceutical Steam Autoclave
Main Parameters
Model | Chamber Size | Overall Size | Net Weight | Steam | Water | Power |
W*H*D mm | W*H*D mm | Kg | Kg/cycle | Kg/cycle | KVA | |
MD-0.8Y | 680*1180*1000 | 2200*2000*1410 | 1480 | 35 | 80 | 3 |
MD-1.0Y | 680*1180*1250 | 2200*2000*1660 | 1630 | 40 | 100 | 4 |
MD-1.2Y | 680*1180*1500 | 2200*2000*1910 | 2030 | 45 | 120 | 4 |
MD-1.5Y | 680*1180*1870 | 2200*2000*2280 | 2230 | 50 | 150 | 4.5 |
MD-2.0Y | 1000*1200*1700 | 2800*2000*2110 | 2490 | 60 | 200 | 4.5 |
MD-3.0Y | 1000*1500*2000 | 2800*2200*2410 | 3050 | 90 | 300 | 5 |
MD-4.0Y | 1000*1500*2670 | 2800*2200*3080 | 3700 | 120 | 400 | 6 |
MD-5.0Y | 1000*1500*3340 | 2800*2200*3750 | 4600 | 150 | 500 | 8 |
MD-6.0Y | 1000*1500*4000 | 2800*2200*4410 | 5400 | 180 | 600 | 8 |
MD-8.0Y | 1000*1500*5340 | 2800*2200*5750 | 7000 | 240 | 800 | 12 |
Application
Medpharm Sterilizers are ideal for Hospitals, Research Laboratory and other Pharmaceutical
Industry.They are designed for safe & efficient sterilization of various items such as Ampoules & Vials, Rubber goods, empty flasks, instruments & utensils, Fabrics wrapped or unwrapped, etc.
Chamber
The chamber is constructed of heavy duty stainless steel 304 to with stand pressure of 2.2 Kg/cm2 corresponding to temperature of 134℃ and also to with stand vacuum of 650-700 mm (26"/28") of Hg in Chamber.Two rails are provided in the chamber bottom for easy loading & unloading of the material on Carriage.
Door
The Sterilizer is supplied with single door or double doors made of 304 S .S. The doors are having automatic sliding type arrangement to open them manually in case of power failure.
The door Gasket will be of Silicon Rubber.
Features
A special provision prevents the operator from opening the door in following events:-
Steam Supply
The Sterilizer is operated by steam supplied from the central boiler.
Steam pressure: 0.4---0.8MPa
Vacuum system
This system is used for effective air removal from Autoclave chamber. It consists of efficient of
water-ring type vacuum pump, condenser etc.
Control system
The control unit is based on advance PLC or Microprocessor technology. This unit control
different sterilization cycles with the help of in -built different program combinations. Control
unit is provided with manual operation facility.
Carriage & Trolley
The loading carriage is made of stainless steel 304. And it has adjustable shelves which
can be positioned according to Height of the goods. It is designed to slide easily from
chamber to trolley.The trolley is made of S.S.lM.S. and its base has four heavy duty castors.
Brochure load down from here
Medpharm medical large steam autoclave brochure.pdf