Versatile and simple operation;
High-visibility color touch screen interface;
Automatic command of cycles;
F0 control or temperature/time of sterilization cycle;
Test cycles for hospital or laboratory loads;
Count of the number of sterilization cycle performed;
Digital printer;
Real-time printing of cycle parameters;
Safety-stop system for door movement in case of obstruction;
Safety system against door opening in mid-cycle in case of power failure;
Incorporated energy and water saving systems;
Extensive list of errors associated to eventual cycle-specific failures;
Equipment fully built in stainless steel, with chamber and doors in AISI 316 stainless steel;
Optionally supplied with 1 or 2 automatic sliding doors;
Control valves pneumatic operated;
Vacuum system by means of a liquid ring vacuum pump;
Accurate PT-100 temperature sensors and electronic absolute pressure sensors;
Integral steam jacket;
Audible and visual Alarms;
Dedicated steam generator with independent water inlet;
Software for documentation and traceability (optional).
Main parameters
Outer size:W120cm*H200cm*D134cm
Steam consumption:25kg/cycle
Water consumption:24kg/cycle
Applications
Medical center CSSD
Competitive Advantage
Vertical sliding door is highly compact and very efficient machine, this Autoclave Steam
Sterilizer is useful in hospitals for the sterilization of rubber goods, textiles, glassware,
containers and various instruments.
Program Parameter
Cycle name | Further details | Application |
Instruments/Tools | 3 times Vacuum, expose temperature 132℃, expose time 8 min, drying time 5 min. | Solid materials to be sterilized: ●Packaged instruments ●Empty glass containers ●Empty metal containers Porous materials to be sterilized: ●Rubber articles |
Clothing | 3 times Vacuum, expose temperature 121℃, expose time 15 min, drying time 25 min. | Materials to be sterilized from which air can’t be easily displaced ●Textile |
User Programmable | All preset parameters can be reset; Multiple customized programs can be set and stored by user. | Customized |
Test programs and utilities | ||
B-D Test | 3 times Vacuum, expose temperature 134℃, expose time 4 min, drying time 3 min. | Test program |
Vacuum Leakage Test | Test the chamber connecting parts’ tightness effect. It’s important to ensure the sterilization effect. | Test program |