sartorius pvc pharmaceutical isolators Single side for Pharmaceutical factory
isolator.pdf brochure
- Hard hatch structure consisting of fully enclosed stainless steel and tempered glass;
- The cabin is welded integrally with 316L stainless steel, and all internal angles of the inner cavity are arc angles, which are convenient for cleaning and sterilization operations;
- Completely unidirectional airflow mode provides an exceptional aseptic operating area;
- The airflow mode is a vertical unidirectional airflow that quickly eliminates biofouling during ventilation;
- The material door is inflated and sealed, with pneumatic locking and working state interlocking function;
- Integrated or independent hydrogen peroxide sterilization can effectively reduce bioactive substance pollution up to 106;
- Integrated YT-601B/YT-608B aseptic detection system to reduce the occupation of operating space;
The isolation system is sealed or has a microbial filtration system (HEPA) that provides air and is self-cleaning. When confined, use only the purified interior surface or use the fast transfer channel for material transfer. When opened, material is only allowed to pass in and out through a specific, designed and verified opening to remove the transfer of contamination. It can be used to isolate active mixtures during aseptic processing or for both sterilization and isolation
- Sterile laminar flow isolators are excellent sterile barriers that provide good environmental control for minimal false positive interference during aseptic processing. With advanced material transfer technology and biological decontamination methods, it can effectively reduce bioactive substance pollution up to 106. Depending on the process characteristics and requirements, special design systems and configurations are available to suit different batch and process flows. Through the configuration selection system determined by the end user, the operating state of the isolator can be changed to a positive pressure operation or a negative pressure operation mode, so that the system has versatility to meet the operational needs of sterile or positive bacteria detection.

Parameter
model | YT-N1200A | YT-N1500A | YT-N1800A |
size | 2000*800*2000 | 2300*800*2000 | 2600*800*2000 |
size of operation | 1200*650*850 | 1500*650*850 | 1800*650*850 |
Air lock chamber size | 750*650*850 |
Power supply | AC220V 50Hz |
Power | 800W |
Operation port | single side 2+1 | single side 3+1 | single side 4+1 |
Operating port size | 240mm diameter |
Airflow mode | engineering turbulence |
Maximum ventilation | 75m3/h |
High efficiency filter | 99.98% H13 |
Pressure control | -99Pa~0Pa |
Operating port protection wind speed | 0.5m / s (refer to ISO 14644-7) |
Hour leak rate | ≤0.5% (refer to ISO 14644-7) |
1. Multi-parameter monitoring system such as cabin temperature, relative humidity and pressure, supporting real-time printing, storage and other functions
2. Built-in weighing device can be built in, and the explosion-proof function can be customized. Comfortable operation, low noise, low operating energy
3. All functions with negative pressure isolators can be installed in Class D clean rooms or rooms without cleanliness


- Sterile laminar flow isolators are excellent sterile barriers that provide good environmental control for minimal false positive interference during aseptic processing. With advanced material transfer technology and biological decontamination methods, it can effectively reduce bioactive substance pollution up to 106. Depending on the process characteristics and requirements, special design systems and configurations are available to suit different batch and process flows. Through the configuration selection system determined by the end user, the operating state of the isolator can be changed to a positive pressure operation or a negative pressure operation mode, so that the system has versatility to meet the operational needs of sterile or positive bacteria detection.