class A class A durable sterility testing isolator manufacturers sterility testing isolator
PVC sterility test isolator.pdf
Configuring a cleaning spray gun (CIP) and drainage system;
An optional PUSH-PUSH filter is available to allow online filter replacement.
The negative pressure isolator is a hard cabin isolator designed for HES requirements. It is mainly used for weighing, batching, unloading, sampling, sampling, etc. of highly toxic and highly sensitizing APIs or preparation products. The negative pressure environment provides optimum protection for operators when weighing and dispensing composite drugs or samples. In addition to the operator protection, the vacuum isolators protect the product and the process.
Used for sampling, weighing and dispensing of highly toxic and highly active pharmaceuticals;
The material transfer and discharge system composed of the air lock chamber and the RPT transfer device can realize the safe transfer of the sample;
The turbulent airflow mode allows for maximum pollution control at low wind speeds, resulting in low energy consumption;
Low negative pressure ensures maximum pollution control while reducing operator fatigue;
Low chamber pressure and low wind speed, combined with a shock-proof weighing platform, provide stable and accurate sample weighing;
The cylinder dust-collecting air filter can be replaced inside the cabin by operating gloves to avoid the harm of pollutants to the human body;
The system integrates the automatic detection of cabin integrity;
Smooth and easy to clean inner and outer surfaces for easy cleaning;
The lighting system is designed outside the isolator cavity for easy maintenance and easy to clean thoroughly in the cavity;
Configuring a cleaning spray gun (CIP) and drainage system;
An optional PUSH-PUSH filter is available to allow online filter replacement....
Parameter
model | YT-N1200A | YT-N1500A | YT-N1800A |
size | 2000*800*2000 | 2300*800*2000 | 2600*800*2000 |
size of operation | 1200*650*850 | 1500*650*850 | 1800*650*850 |
Air lock chamber size | 750*650*850 | ||
Power supply | AC220V 50Hz | ||
Power | 800W | ||
Operation port | single side 2+1 | single side 3+1 | single side 4+1 |
Operating port size | 240mm diameter | ||
Airflow mode | engineering turbulence | ||
Maximum ventilation | 75m3/h | ||
High efficiency filter | 99.98% H13 | ||
Pressure control | -99Pa~0Pa | ||
Operating port protection wind speed | 0.5m / s (refer to ISO 14644-7) | ||
Hour leak rate | ≤0.5% (refer to ISO 14644-7) |
WINTEAM's new generation of intelligent soft cabin sterility inspection isolator, which is the fourth generation of the isolation products of Tailin. It has been upgraded from the appearance of the product or from the technical aspects of the product. The system is more powerful and laminar. The airlock transmission compartment is perfectly combined with the experimental cabin to realize the maximum operation space and the need for rapid sterilization and delivery of materials. It fully meets the USP/EP/CHP sterility test regulations and has stable performance; while ensuring the quality of the drug inspection, it is greatly Improve the efficiency of inspection work:
1. To achieve the maximum operation space and the need for rapid material transfer, fully meet regulatory requirements and stable performance.
2. The operating space maximizes the design while having a more compact form factor (reduced height and width) for easy installation
3. Intelligent control system, built-in equipment automation operation and management program, with multiple expansion ports and applications, powerful
In addition, there are are operating videos for our products for the convenience of customers to use.