SPECIFICATIONS:
- The function of oxygen:
1. Supplying oxygen to patients can assist in the treatment of heart
head blood-vessel, respiratory system, such as chronic obstructive
pneumonia disease, rehabilitation and hypoxia conditions.
2. Oxygen could improve body function, especially suit for elderly,
pregnant women, physique, students etc. Who got physiological
hypoxia. It may also reduce fatigue after heavy manual or mental
consumption.
THE SCOPE OF PRODUCTS
1. For hospital and homecare to supply oxygen to patients.
STRUCTURE
1. The product is mainly composed of oxygen making host, flow meter, humidifier.
2. Safety plastic shell.
3. With timing power off function.
4. Power loss alarm function.
5. The compressor with thermal protector guarantee better security.
6. With nebulizer function. (7F-3W)
FEATURES
SPECIFICATIONS
1. Maximum recommended flow: 3L/min
2. Flow Range: 0.5 ~ 3L/min
3. Change in maximum recommended flow when back pressure of
7kPa is applied: < 0.5L/min
4. Oxygen Concentration: 93% ± 3%
5. Maximum Outlet Pressure: 45kPa ± 4.5kPa(6.52psi ± 0.65psi),
6. Pressure Relief Mechanism Operational at: 250kPa 土 25kPa
(3625psi±3.63psi)
7. Noise: w55dB(A)
8. Maximum Nebulization Rate: ^0.15mL/min
9. Power Supply:
AC120V±10% AC220V±10% AC230V±10%
50Hz 60HZ 1.4 amps(AC220V-240V); 2.9 amps(AC120V)
10. Input Power: 320VA
11. Weight: 24kg
12. Dimension: H55.6cm (21.9") x W28.5cm (1123”) x D47cm (18.52")
13. Minimum operating time: 30minutes
14. Altitude: Up to 1828 meters (6000 ft) above sea level without
degradation of concentration levels. From 1828 meters (6000 ft) to
4000meters (13129 ft) below 90% efficiency.
15. Minimum Operating Time: 30minutes
16. Electric classification: Class II, type BF application part
17. Work system: work continuously
18. NORMAL OPERATION CONDITIONS:
Temperature range: 10°C ~ 35°C (50°F ~ 95°F);
Relative humidity: 30%-75%
Atmosphere pressure: 860hPa ~ 1060hPa(12.47psi ~ 15.37psi);
NOTE: When storage temperature is below 5°C, please keep the
equipment in normal working condition for at least 4 hours before
using.
19. Temperature at oxygen outlet: <46°C
20. Suggestion: The length of nasal oxygen cannula should be less
than 15.2 meters and should not be pinched.
21. Transportation and storage conditions:
Temperature range: -20°C ~ +60°C(-68°F ~ +140°F)
Relative humidity: 10%-93%, no condensation
Atmosphere pressure: 700hPa~ 1060hPa(l 0.15psi * 15.37psi).
NOTE:Oxygen concentrator should be stored with out intense
sunlight no corrosive gas and ina well ventilated room, avoid jolt
and inverted recumbent transport.