Display Accuracy 0.1 °c Environmental Laboratory High Temperature Test Chamber
High temperature test chamber JD-8002-200
| |
Box knot Structure size |
1, inner size: 450 (width) × 400 (deep) × 400 (height) mm
2, the outer dimensions are about: 720 (width) × 590 (deep) × 1020 (height) mm + 110 casters, whichever is the real thing;
3. Operation panel: The control panel is located above the machine table;
4, open the door: single door from right to left;
5, can be window: the door with a window, specifications W200 * H250mm;
6, the inner material is: 430 # mirror steel plate, thickness 1.0mm;
7, the outer material is: cold-rolled steel plate, thickness 1.0mm. Powder paint treatment (bei white);
8, compartment: inside the second layer, with two stencils;
9, the insulation material is: glass fiber rock wool, good insulation effect;
10. The sealing material is: high temperature resistant foamed silicone strip;
11. The machine is equipped with brake movable casters;
12. The overall structure is floor-standing.
|
Temperature control system |
1. The temperature control is PID digital display instrument (Taiwan Taisong EM705), single-point temperature control, automatic calculation, PV/SV simultaneous display, button setting;
2, timing function: 1-9999 (optional setting hours, minutes, seconds) temperature to the timing, time to stop heating power supply, and alarm prompt;
3. Temperature sensing line (K) type;
4, the output is 3-32V;
5. The current controller is: SSR solid state relay;
6, the heating material is: high temperature heating tube;
7, temperature range: room temperature +20 ~ 200 °C temperature adjustable:
8, heating rate: 100 ° C about 15 minutes;
9, control accuracy: ± 0.5 ° C;
10, display accuracy: 0.1 ° C;
11, temperature deviation: RT + 20 ~ 200 ° C ± 2.5 ° C. (no-load test)
|
Air supply system |
1. Air supply mode: internal horizontal air supply circulation;
2, the motor is a special type of temperature resistance, motor power 370W / 50Hz;
3. The fan is 7.5-inch multi-blade turbine type;
4. The air inlet is on the right side of the machine and the air outlet is on the left side of the machine.
|
power supply Device |
1, voltage AC 220V/50Hz single phase
2, current 16A
3, power 3.5KW
|
Security system |
1. Over-temperature protection system: When the temperature is out of control, when the over-temperature protection set temperature is exceeded, the heating supply is automatically stopped, thus protecting the safety of products and machines;
2. Circuit protection system: grounding protection, fast insurance, etc.
|