Human Medical Temperature Sensor Skin Disposable 2.252K 10K
Brief Instruction:
Skin disposable temperature probe medical temperature sensor is widely used for human body surface temperature measurement. The sensor probe is epoxy and metal disk—the rated resistance value is 2.252KΩ or 10KΩ. The total length is 0.7&0.9Meters, and the sensor measuring tolerance is within ±0.1℃. Used for monitors, suitable for YSI400 series products replacement.
Features:
DisposableTolerance(25~45)℃±0.1℃
Operating temperature range: 0℃~50℃
Application:
Applicable to Medical monitors
Structure and Dimensions:(Unit: mm)
NO. | Name | Material Specification | QTY. | Remarks |
1 | Chip | NTC Thermal Chip | 1 | / |
2 | Cable | PVC 28AWG (7/0.1) Cable | 1 | white |
3 | Probe | Epoxy resin and Stainless steel housing | 1 | / |
Electrical Specification:
Item | Code | Test condition | Performance requirements | Unit |
(Rated resistance) | R25℃ | +25℃±0.02℃ | 2.252 | kΩ |
(B value) | B25/50 | +25℃±0.02℃, +50℃±0.02℃ | 3935 | K |
(Response time) | τ | 23℃(air)→44℃(水/water),≤±0.1℃ | ≤150 | S |
(Max.Energy Dissipation) | Pmax | In water(≤1.2mA),≤0.02℃ | Approx 50 | μW |
(Withstand voltage) | Hi-pot | 静止空气中 AC750V 1mA 1s In still air | No break down | V |
(Operating Temp. Range) | OTR | In still air | 0~50 | ℃ |
Standard packing (inner packing): 1 pcs/polybag
Operation Notice:
1. Application: Temperature measure ;
2. The current flowing through the NTC is controlled within the Pmax range to avoid the measurement error caused by the self-heating of the component.
3. Storage Temperature is -10℃~40℃,≤75%RH; Avoid a corrosive gas and a sunlit environment; After opening the package, it needs to be sealed again.