Chemical Environment Temperature And Humidity Climate Box AC 220V
Exported temperature and humidity climate chamber Application:
Exported temperature and humidity climate chamber is also called environment testing machine, used to test various types of materials for their tolerances of heat, cold, dry, humidity. It is applied for quality inspection for all types of electronic products, electric appliance, and products from communication, instrument, automobile, plastic, metal, food, chemical, building materials, medical and aerospace industries.
Working Principle:
This is so called Fast Shipping Thermal Shock Test Chamber, Thermal Chamber, Thermal Shock Chamber.
Sample is placed in the test chamber, more extreme temperature than test temperature can be set in high-temperature chamber and low-temperature chamber. When it comes to low temperature test, cold chamber door open, and low-temperature chamber working together with test chamber.
When converted into high-temperature test, cold chamber door closed, hot chamber door are opened, and test chamber working together with high-temperature chamber.
Conversion of mechanical action (transferring from high temperature to low temperature or low temperature to high temperature) can be completed in less than 1second, and the temperature can be quickly stabilized. During the whole test, test sample is no need to be moved, and without any human intervention.
Specification:
Model | GSKA80 | GSKA150 | GSKA220 | GSKA800 | GSKA1000 | |
Internal Dimension WxHxD (mm) | 400x500x400 | 500x600x500 | 600x750x500 | 1000x1000x800 | 1000x1000x1000 | |
Internal Dimension WxHxD (mm) | 950x1650x950 | 1050x1750x1050 | 1200x1900x1150 | 1600x2000x1450 | 1600x2100x1450 | |
Temperature Range | Low Temperature(A:25ºC B:0ºC C:-20ºC D:-40ºC E:-60ºC F:-70ºC) High Temperature 150ºC | |||||
Humidity Range | 20%~98%R.H.(10%-98% R.H. / 5%-98% R.H ,is optiona , need Dehumidifier) | |||||
Indication relolution/Distribution uniformity of temperature and humidity | 0.1ºC; 0.1% R.H/±2.0ºC; ±3.0% R.H. | |||||
Control Accuracy of Temperature and Humidity | ±0.5ºC; ±2.5% R.H. | |||||
Temperature Rising/ Falling Velocity | Temperature rising approx. 0.1~3.0ºC/min ; temperature falling approx. 0.1~1.5ºC/min; ( Falling Min.1.5ºC/min is optional) | |||||
Inner and Outer Material | Interior material is SUS 304# stainless steel, exterior is stainless steel or SEE cold-rolled steel with paint coated. | |||||
Insulation Material | Resistant to high temperature, high density, formate chlorine, ethyl acetum foam insulation materials | |||||
Cooling System | Wind cooling or water cooling, (single segment compressor-40ºC, double segment compressor -70ºC) | |||||
Protection Devices | Fuse-free switch, overloading protection switch for compressor, high and low voltage coolant protection switch, over-humidity and over-temperature protection switch, fuses, fault warning system, water short storage warning protection | |||||
Optional Accessories | Inner door with operation hole , Recorder, Water Purifier, Dehumidifier | |||||
Compressor | French Tecumseh Brand, Germany Bizer Brand | |||||
Power | AC220V 1∮ 3 lines, 50/60HZ , AC380V 3∮ 5 lines , 50/60HZ | |||||
Approx. Weight (Kg) | 150 | 180 | 250 | 400 | 450 |
We keep good quality and competitive price to ensure our customers benefit ; We respect every customer as our friend and we sincerely do business and make friends with them,no matter where they come from. Each batch of parts is done quality sampling inspection before into the warehouse, and Each product before delivery is tested by the national authoritative third-party inspection agency