Application: Dry State Lint Tester to test the coefficient of linting and other particle s generation of surgical drapes, gowns and clean air suits for patients, clinical staff and equipment in the drying state.
Standards: ISO 9073-10, INDA IST 160.1, DIN EN 13795-2, YY/T 0506.4 |
Model | GT-RA16 | |
Diameter of the round plate | Φ82.8㎜ | |
Reciprocating frequency | 60times/ min | |
Rotating Angle | 180°/time (clockwise & anticlockwise alternating) | |
Max start distance of 2 round plates | (188±2)㎜ | |
Straight stroke of the dynamic plates | (120±2)㎜ | |
Size of anti-static electricity test chamber | 300mm×300mm×300mm | |
Diameter of air collector sampling end | (40±5)㎜ | |
Laser dust particle counter | Measure range of particle size | 0.3μm~25μm |
8 Test channel | 0.3μm, 0.5μm, 0.7μm, 1.0μm, 2.0μm, 3.0μm, 5.0μm, 10.0μm | |
Air flow speed | (28.3±1.4)L/min | |
Sampling period | 1~59min59s optional | |
Print function | Insert printer, can print counting result | |
Max collector concentration | 35,000 particles/L | |
Power supply | AC 220V 50Hz |