Automatic Shutdown Digital Torque Meter With Data Storage Digital torque tester
Features:
Automatic shutdown (after 10 minutes to place self-closing (OFF))
Data storage function (which can record 100 torque value, with storage, browsing, some of the deleted, delete all the features.)
Calculate the maximum (MAX), minimum (MIN), average (AVE).
Auto-delete function of the value of torque (with automatic reset switch adjustable time required)
Auto-zero (that can automatically return to zero)
Can be set torque value. Buzzer and can be prompted operators
Specifications
Model No. |
TP-1N |
TP-5N |
TP-10N |
TP-20N |
Test range |
0.015-1.000N.m
0.15-9.00Lbf.in
0.15-10.00Kgf.cm
|
0.15-5.00N.m
1.5-44.0Lbf.in
1.5-50.0Kgf.cm
|
0.15-10.00N.m
1.5-90.0Lbf.in
1.5-100.0Kgf.cm
|
0.50-20.00N.m
5.0-180.0 Lbf.in
5.0-200.0Kgf.cm
|
Accuracy |
±0.5%(F.S.) |
Size(mm) |
200*124*65.2(H) |
Weight |
1.8kg |
Full time use of electricity in a row |
8 Hours |
Battery life |
300 times |
Special charger |
Input Voltage:AC100V-240V 50/60Hz
Output Voltage:DC7.3V - 8V 0.5A
|
Operating instructions:
First check the state of the battery charge. When the power is on and the battery has an insufficient
charge,the display reads"LOBAT. "When this happens, use the battery charger to recharge the battery
for at least 3 hours and no more than 6 hours.
1. Make sure the digital torque meter is stable.
2. Conduct zero adjustment. Set [ MODE ] switch to "TRACK", then press both the【SET】and [▲】 switches at the same time for less than 2 seconds. This adjusts the meter to zero.
(Note: If you press the switches for than 2 seconds the meter will go into another mode. If this happens, press【 RESET】, then repeat the zero adjustment process.
3.Set【MODE】 switch to"PEAK"or" F. PEAK"
In peak mode the peak torque value reached during measurement is displayed. Thus, the display
does not change Model TP-10N measures up to10 Nm of torque, which corresponds to a display value of 15. Please conduct measurements within this range.
In "F. PEAK" mode the first peak value is displayed and does not change even if higher torque is applied
4. Press the【 RESET】 switch to cancel the display.
5. Minus signs appear in front of measurements of counterclockwise torque.
6. To output data using a Mitutoyo Input Tool, set the No. 3 connector of the Dipswitch(page 6) to the on position(so the "P. ON"LED lights),set the【MODE】 switch to"PEAK" and press the【 RESET】 switch( which functions the same as the Enter key on a computer). The Input Tool will convert the measurement data for direct input to a spreadsheet application. In order to output data, please conduct measurement within the Instrument's peak measurement range, as indicated in the specifications.
Minus signs denoting counterclockwise torque are transferred with the measurement data.
7. Use the Fidaptor to measure torque of power screwdrivers. as expiained below.
8. It is possible to fashion adapters for torque measurement of other types of tools. ( See"Using the
adapter socket for torque measurement of instruments other than screwdrivers. "
9. Be sure to turn off the power and remove the tool from the socket when you are finished measuring.