Time-resolved Fluorescence Immunoassay Analyzer, testing HbA1c/tumor markers/COVID/hCG...
The Immunofluorescence analyzer is an instrument for detecting the concentration of corresponding test items by human blood, urine and feces, such as diabetes, inflammation, cardiac markers, coagulation, fertility, tumor markers, COVID-19 and others. It is usually use in clinics, hospitals, small labs, pharmacies and other medical institutions.
To meet different test items, we should choose the corresponding detection ID chip and test card, then drip the sample liquid to the test area on the test card. Lay the test card flat on the table.
When we are ready to start the test, connect the power to the outlet and startup, then insert the ID chip to the chip port, the system can recognize the type of detection according to the ID chip. After then, insert the corresponding test card to the card port. If the test card were already reacted completely, we can choose the quick test mode by finger touch on the screen. Otherwise, if the test card were not reacted complete, we can choose the standard test mode.
Quick test mode needs to calculate the reaction time of test card by oneself, it can test multiple samples in a short time. Standard test mode just can test sample one by one, and the instrument will calculate the reaction time automatically.
After chose a test mode, click the start test button on the screen, then the excitation light will irradiate to the test card and reflect the emitted light to the signal connection. In this way, we can check out the quantity concentration of sample and display on the screen. We can get the result between 3-15 min according to different test item. Print the result with the printer on the instrument.
To meet the demands of different applications, we have three type of product models, you can choose according to your requests. CHF-600 has 6 channels, CHF-200 and CHF-100 has 1 channel.
No. | Name | Quantity |
1 |
Immunofluorescence Analyzer
|
1 |
2 |
Power adapter for instrument (including the power cord)
|
1 |
3 |
Printer paper
|
1 |
4 |
Operation manual
|
1 |
5 |
Quality Control Card
|
1 |
6 |
Certificate of qualification
|
1 |
7 |
Packing list
|
1 |
8 |
Warranty
|
1 |
Excitation light source
|
LED
|
Excitation spectrum
|
Central wavelength λ0=365nm
|
Absorption spectrum
|
Central wavelength λ1=610nm
|
Test Specimen
|
Whole Blood, Serum plasma, urine, feces, quality control solution
|
Operation environment
|
Android 5.1.1
|
Interface
|
Serial port,USB,Ethernet network
|
Built-in Printer
|
Thermal printer
|
Size
|
280X 280 X 170 mm ±5mm
|
Weight
|
2Kg ±100g
|
Repeatability
|
CV≤15%
|
Stability
|
CV≤15%
|
Operation Environment:
Indoor use
|
|
Ambient temperature
|
5℃~40℃ |
Relative humidity
|
10%-80%(non-condensing) |
Atmospheric pressure
|
860hPa~1060hPa |
Altitude
|
Up to 2000m |
pollution degree
|
2 |
Voltage
|
100-240VAC±10% |
Power frequency
|
50/60Hz |
Input power
|
60VA
|
Storage Environment:
Temperature
|
-10℃~+50℃
|
Humidity
|
<85%
|
Performance:
Precision
|
CV≤15%
|
Accuracy
|
Relative Deviation≤15%
|