2 Wire RTD Multifunction Process Calibrator Thermocouple Calibration Equipment
RTD Source Resistance and Thermistance Calibrator Multifunction Source Signal Calibrator
Output function and technical index
Output function | Range | Output range | Resolution | Accuracy | Illustration | |
Simulate resistance OHM | 400Ω | 0.0~400.0Ω | 0.1Ω | 0.05%+2 | Incentive current is set as:±0.5~±3mA When the incentive current is set as ±0.1~0.5mA, add an extra 0.1Ω to additional error. The accuracy does not include lead resistance | |
4000Ω | 0~4000Ω | 1Ω | 0.05%+2 | Incentive current is set as ±0.05~±0.3mA The accuracy does not include lead resistance | ||
Thermal resistance RTD | Cu10 | -10°C~250°C | 0.1°C | 0.05%+6 | Incentive current is set as ±0.5~±3mA When the incentive current is set as ±0.1~0.5mA, add an extra 0.5°C to additional error. | Employs Pt(385) standard temperature The accuracy does not include lead resistance |
Cu50 | -50.0°C~150.0°C | |||||
Pt10 385 | -200.0°C~850.0°C | |||||
Pt100 385 | -200.0°C~850.0°C | |||||
Pt200 385 | -200°C~630°C | Incentive current is set as± 0.05~±0.3mA
| ||||
Pt500 385 | -200°C~630°C | |||||
Pt1000 385 | -200.0°C~630.0°C |
Output function and technical index
Output function | Range | Output range | Resolution | Accuracy | Illustration |
Resistance OHM | 500Ω | 0.0~500.0Ω | 0.1Ω | 0.05%+2 | Measurement current:about 1mA Open circuit voltage: about 2.5V The accuracy does not include lead resistance |
5000Ω | 0~5000Ω | 1Ω | 0.05%+2 | Measurement current:about 1mA Open circuit voltage: about 2.5V the accuracy does not include lead resistance | |
Thermal resistance RTD | Cu10 | -10.0°C~250.0°C | 0.1℃
| 0.05%+0.6℃
| The incentive current is set as:±0.5~±3mA When the incentive current is set as ±0.1~0.5mA, add an extra 0.5°C to additional error. |
Cu50 | -50.0°C~150.0°C | ||||
Pt10 385 | -200.0°C~850.0°C | ||||
Pt100 385 | -200.0°C~850.0°C | ||||
Pt200 385 | -200°C~630°C | Incentive current is set as ±0.05~±0.3mA
| |||
Pt500 385 | -200°C~630°C | ||||
Pt1000 385 | -200.0°C~630.0°C |
General Feature
| : | two 1.5V alkaline batteries(LR6) |
l power consumption | : | about 70mA/3V |
| : | 30V(within terminals or between terminal and earth ground) |
| : | 0°C~50°C |
| : | ≤ 80%RH |
| : | ≤ - 10°C~55°C |
| : | ≤ 90%RH |
| : | 0.1×(dedicated accuracy)%/℃(5℃~18℃,28℃~40℃) |
| : | 180(L)×90(W)×47(D)mm(with protector) |
| : | about 500g |
| : | User’s Manual, industrial testing lead H000001 (including alligator clip) |
| : | complies with IEC1010 (safety standard issued by International Electrician Committee) |