Wayeal IC6200 Series Integrated Anion or Cation Ion Chromatography Anlayzer System
Introduction of Ion Chromatography Instrument
Ion chromatography (or ion-exchange chromatography) is a chromatography process that separates ions and polar molecules based on their affinity to the ion exchanger, based on the electrical charge of the protein, which has both positively and negatively charged chemical groups and can be altered by changes in the pH of the buffer, utilizing a gradient of linearly increasing salt concentration.
Features of Ion Chromatography Instrument
1. Eluent Bottle
Sealed receiver design
2L thick-walled polypropylene eluent bottle;Double bottle design with additional deionized water for rinsing.
2. In-line Degasser
Teflon@AF tube is used to shorten the eluent change timeTwo channels for simultaneous in-line degassing of eluent and rinse liquids.
3. Precision PEEK Pump
Full PEEK pump head is resistant to acid, alkali and reversed-phase organic solvents.avoiding metal contamination.Two-stage suspension drive technology, extending the service life of the seal.Tandem dual plunger mode with electronic pulsation suppression technology provideshigh precision, low pulsation and low drift eluent;With high and low pressure protection function
4. Solvent Switching Device
AlI PEEK material allows for seamless switching between eluent and rinse fluids andonline replacement without stopping the pump;One-touch flushing" and "one-touch maintenance" function with software control.
Parameters of Ion Chromatography Instrument
Wayeal IC6200 Series Integrated Anion Or Cation Ion Chromatography System | |
System Parameter | |
Qualitative Repeatability | ≤0.5% |
Quantitative Repeatability | ≤1% |
Pump System | |
Flow Rate | (0.001~10.000)mL/min |
Pump Head Pressure Resistance | (0 ~ 35) MPa |
Flow Rate Setting Inaccuracy | <0.1% |
Flow Stability | <0.1% |
Pressure Display Accuracy | 0.01Mpa |
Pressure Pulsation | <0.5% |
Conductivity Detector | |
Baseline Noise | ≤0.1%FS |
Baseline Drift | ≤0.5%FS |
Minimum Detectable Concentration (Cl) | ≤0.0002μg/mL |
Minimum Detection Concentration (Li) | ≤0.0001μg/mL |
Conductivity Cell Volume | ≤ 0.6μL |
Conductivity Detection Range | 0-15000μS/cm |
Detector Resolution | 0.00238nS/cm |
Conductivity Cell Temperature Setting Error | ±0.01°C |
Resolution of Conductivity Cell Temperature Display | 0.001°C |
Suppressor | |
Constant Current Source Range | 0-500mA in 0.1mA increments |
Column Oven | |
Temperature Range | Room temperature +5°C~85°C |
Control Temperature Stability | ≤ 0.1°C |
Autosampler | |
Number of Sample Vials | ≥ 120 (1.5mL standard sample bottle) |
Repeatability | Full loop injection < 0.3% RSD, Partial injection < 0.5% RSD, Non-destructive injection < 1 % RSD (injection volume > 5uL) |
Syringe Size | Standard: 500μL; optional; 250μL, 1000μL, 2500μL |
Loop Size | Standard: 100μL; optional: 20μL, 50μL, 200μL |
Linearity | >0.999 |
Cross-contamination | <0.01% |
Eluent Generator | |
Flow Rate Range | 0.001 to 3.000ml/min |
Eluent Concentration Range | KOH, MSA-0.1 to 100mM |
Concentration Accuracy | 0.1mM |
Original Concentration of Electrolyte | 25% KOH/MSA |
Volume of Electrolyte | 1000mL |
Maximum Working Pressure | 21MPa (3000psi) |
Gradient Accuracy | 1.5% |
Industry Applications of Ion Chromatography Instrument
Some typical applications of ion chromatography include: