Zero Auto Tracking Dc 24V Guide Rail Weighing / Force Measuring Led Control With Rs232
Main Menu | Second Menu | |||
F6.FAC | Factory Adjustment | Special for manufacturer. | ||
-SPAn- | 【mV】Exfactory Span Adjustment: Use standard weighing test equipment to adjust the weighing signal interface for normalizing Span Coefficient to 1. | |||
-AdtS- | Weighing Signal Linearity Test. | |||
-AVZF- | AO [0~20mA] Zero/Full Adjustment. VO [0~10V] Zero/Full Adjustment. | |||
-AVtS- | AO [0~20mA] Linearity Test. VO [0~10V] Linearity Test. | |||
-dIdo- | DI/DO Input&Output Test. | |||
-dEFU- | RAM Reset: Reset to factory defaults. | |||
-dStS- | Display/DO Reliability Test. | |||
F7.InF | Product Information | Only for query. | ||
--VEr- | Version No. | |||
--Sn-- | Serial No. | |||
-dAtE- | Exfactory Date. | |||
-HArd- | Hardware Configuration. |
Dc 24V Guide Rail Weighing / Force Measuring Led Control Module With Rs232 and Rs485
Main Features:
Keypad Operation
Key Name | Function | Description |
【■】 | 【MENU】 | Enter Main Menu / Exit. |
【LOCK】* | Key-locking / Key-unlocking. | |
【►】 | 【►】 | Cursor shifts right. |
【ZERO】* | Manual Zero without Power-down Protection. | |
【▲】 | 【▲】 | Display the next interface or option. |
【SET】* | Setpoint Parameters Setting. | |
【8】 | 【ENT】 | Enter / Save / Clear Alarm. |
【CLS】* | Clear Screen [Clear the display value of Peak Hold Value and Weight-checking Value]. |
User Interface Diagram
Installation
Main Menu
Main Menu | Second Menu | ||
Sign | Function | Sign | Description |
F1.SEt | Parameter Setting | -SCAL | Scale parameters setting. |
-CALP- | Calibration parameters setting. | ||
-SEtP- | Setpoint parameters setting. | ||
-SErP- | Communication parameters setting. | ||
-dISP- | User interface parameters setting. | ||
F2.CAL | System Calibration | -ZEro- | Zero Calibration without loading on the weigher to correct Zero Value. |
-dAtA- | Data Calibration: Input the specification parameter values of loadcell according to the actual configuration of the weighing system to correct Span Coefficient. If there is no access to get the specification parameter values for Data Calibration, then it’s necessary to do Load Calibration. | ||
-LoAd- | Load Calibration: After doing Data Calibration, if there are conditions for Load Calibration, do Load Calibration with loading standard weight on the weigher to correct Span Coefficient further for higher weighing accuracy. | ||
-SEgC- | Segmenting Span Correction: After doing Zero Calibration and Load Calibration [or Data Calibration], it’s allowed to do Segmenting Span Correction with loading standard weight on the weigher by 10 Correction Points to get Span Correction Ratio of 10 linearity segments for higher weighing accuracy. | ||
F3.rEC | Unused. | ||
F4.CLn | Clear Screen: Clear the display value of Peak Hold Value and Weight-checking Value. | ||
F5.Loc | Key-locker | -oPEn- | Key-unlocking. |
-Locc- | Key-locking. | ||
-PASS- | Password Set.
|